Overview of Memosens CPS96E
The Memosens CPS96E by Endress+Hauser is a high-performance, digital combined sensor designed for simultaneous pH, ORP (redox), and rH (in rH mode) measurements
Key Specifications
- Measurement Ranges:
- ORP: –1,500 mV to +1,500 mV
- pH: 0 to 14
- Process Temperature: 0 – 110 °C (32 – 230 °F)
- Process Pressure: 0.8 – 14 bar absolute (approx. 11.6 – 203 psi)
- Ingress Protection: IP68 rated for robust environmental sealing
- Sensor Materials:
- Shaft: Glass suited to processes
- pH membrane: Type B glass
- Reference system: Ag/AgCl with stabilized gel and ion trap
- ORP electrode: Platinum disc
- Seals: FKM O-rings, PPS fiber-glass reinforced coupling, ceramic metal oxide nameplate
Distinctive Features & Advantages
- Open-Aperture Design: Reduces fouling and avoids diffusion potential build-up—ideal for media with high levels of suspended solids such as dispersions, emulsions, and precipitates
- Memosens 2.0 Technology:
- Digital, inductive signal transmission (non-contact) ensures reliable, moisture-resistant data transfer
- Supports lab calibration and stores extensive calibration and process data, enabling trend analysis and predictive maintenance—perfect for IIoT integration
- Sensor Quality Monitoring: Includes a platinum electrode to monitor reference impedance continuously, helping detect deterioration or poisoning
- Robust & Poisoning-Resistant: Stabilized gel with ion trap protects reference system from contamination
- High Process Safety: Digital inductive coupling removes the need for direct electrical contacts, enhancing safety in hazardous environments
- Certification for Hazardous Zones: ATEX, IECEx, CSA C/US, NEPSI, Japan-Ex, and INMETRO approvals—suitable for Zones 0, 1, and 2
Typical Applications
The CPS96E is engineered for challenging, contaminated, or solid-laden media, making it a go-to for:
- Chemical processes
- Pulp and paper industry
- Paint and flue gas cleaning
- Media with dispersions, precipitations, emulsions, or general contamination
Summary Table
Feature
|
Description
|
Measurement Capabilities
|
Simultaneous pH, ORP, rH (in rH mode)
|
Design
|
Open aperture, stabilized gel with ion trap, platinum reference electrode
|
Data & Connectivity
|
Digital signal via Memosens 2.0, lab calibration, data logging
|
Durability & Safety
|
IP68, inductive coupling, hazardous-area certifications
|
Optimal Use Cases
|
Suspended solids, emulsions, contaminated or high-fouling industrial media
|